What to do if you encounter a small fault with a peristaltic pump?
If the peristaltic pump does not respond when powered on, then please check in the following order. If the pump still does not respond after checking, you can contact our technical support; if the circuit protection device is installed, you need to confirm that the circuit is not tripped, confirm that the power plug is inserted into a normal working socket, check whether the power cord is firmly inserted, and confirm whether the power interface fuse is blown
If the pump is on but not running, first you can check if the coupling is damaged, if not, cut off the power, turn the pump head manually and toggle the roller to see if it is rotating properly.
The pump transmits a low or no flow rate, give priority to whether the flow rate is set too low, for example when setting 0.1ml the roller barely moves, set a higher flow rate when testing. Next, you can also follow these steps to check. Check that the reagents are normal, check that the pipes are not tangled or blocked, check that all valves are open, check that the hose is in the middle of the rotor, check that the hose is not cracked or damaged, check the running direction, check that the pump head roller can be turned flexibly
The external control cannot be controlled, regardless of the digital display type and LCD type, you can see in the display whether the external control icon has been switched successfully, such as has been successful, you can check whether the external control line is correctly connected, the signal source settings are normal.
The above is a short answer to the solution of the failure, if still ca not solve it, you can also contact us.