🧲 Thursday's MRI Discussion 🧲
🤔 What is Partial Fourier in SIEMENS MRI?

📝Phase Partial Fourier in SIEMENS MRI (also called Partial Phase, Half Fourier, or Asymmetric Sampling) is a technique where the scanner acquires less than 100% of phase-encoding lines in k-space and mathematically reconstructs the missing part.
---

✅ Simple Definition

Phase Partial Fourier = acquiring 60–80% of phase-encoding data and estimating the rest using symmetry of k-space.

It reduces scan time but may slightly reduce SNR and increase image noise if too much is skipped.
---
🔍 How it works

k-space is symmetric about its center.

SIEMENS collects only the low-frequency + some high-frequency lines.

The missing high-frequency phase lines are reconstructed using:

Homodyne detection

POCS (Projection Onto Convex Sets)
---
📌 Where Phase Partial Fourier is used

Used in almost all sequences:

T1, T2 spin echo

Fast GRE sequences

3D sequences

DWI/EPI
---
🎯 Benefits

Benefit Explanation

Shorter scan time Fewer phase lines → faster TR loops
Less motion artifact Shorter time reduces patient movement
Useful for thin slices & 3D scans Improves feasibility
---
⚠️ Disadvantages

Issue Why it happens

Slight SNR loss Less data acquired
Phase artifacts If too much reconstruction needed
Ghosting In EPI if PF too high
---
💡 SIEMENS Settings You’ll See

In Siemens protocol, Phase PF appears as:

Partial Fourier (Phase): 6/8, 7/8, 5/8

Sometimes written as PF = 0.75 or 0.625

Common setting:
👉 6/8 (75%)
giving best balance of speed and image quality.
---
📘 Practical Example

If phase matrix = 320
Partial Fourier = 6/8

Scanner actually acquires: 320 × 0.75 = 240 lines

Remaining 80 lines = reconstructed.
---
🧠 Quick Summary (Easy to Remember)

Phase Partial Fourier = 60–80% phase data.

Purpose = Faster scan + less motion.

Cost = Slight SNR drop + possible artifacts.
---

12/12/2025
10/12/2025
10/12/2025

Address

Romsey Road
Southampton

Telephone

0712610085

Website

Alerts

Be the first to know and let us send you an email when Department Of Medical Imaging posts news and promotions. Your email address will not be used for any other purpose, and you can unsubscribe at any time.

Contact The Practice

Send a message to Department Of Medical Imaging:

Share

Share on Facebook Share on Twitter Share on LinkedIn
Share on Pinterest Share on Reddit Share via Email
Share on WhatsApp Share on Instagram Share on Telegram