GE Alpha RT PMS using phantom and weight scale to check compression.
21/05/2025

GE Alpha RT PMS using phantom and weight scale to check compression.

18/04/2025

Philips Bucky Diagnost at Capitol Medical running 24/7 for over 2 1/2 decades. Very reliable with minimum down time. Common faults in the collimator: timer board and tube stand switches.

GE TMX common B4 charger IGBT failures. The IGBT collector PCB trace is very close above the gate - emitter resistor thu...
19/03/2025

GE TMX common B4 charger IGBT failures. The IGBT collector PCB trace is very close above the gate - emitter resistor thus when presence of dirt - moisture that leads to small arcing.
Best is to put mylar insulation as shown below

07/12/2024

Rare case insulating oil gets inside the xray tube insert.

Mammography phantom innovation using sewing thread, rice grains, ziplock, and faucet filters.
15/08/2024

Mammography phantom innovation using sewing thread, rice grains, ziplock, and faucet filters.

D100 problem on Axiom Aristos FX Plus at Gensa. No communications with XCU
28/04/2024

D100 problem on Axiom Aristos FX Plus at Gensa. No communications with XCU

GE Prestige S1 Tube Angulation Controller PCB with a burned SCR screw /nut that damaged the longitudinal follower-pot at...
27/04/2024

GE Prestige S1 Tube Angulation Controller PCB with a burned SCR screw /nut that damaged the longitudinal follower-pot at Gensa

Quantum Large Filament transformer problem and removal
23/04/2024

Quantum Large Filament transformer problem and removal

09/04/2024

BMI Jolly Monoblock Static Testing

GE RT Alhpa Mammo with burned HV inverter assembly.
13/03/2024

GE RT Alhpa Mammo with burned HV inverter assembly.

Angell kV board rebuilt. Unit was down for over a year.
13/03/2024

Angell kV board rebuilt. Unit was down for over a year.

Initial report that there was an accident that caused damages to both the RCB and Main Contactor and was replaced by the...
13/03/2024

Initial report that there was an accident that caused damages to both the RCB and Main Contactor and was replaced by the previous service provider.
It was said that the original problem was a Communication Problem between console and generator thus 2 MCB were orderd to resolve problem this.
During our trouble shooting we found that the +24VAA got shorted by the Enode Board and that caused damage to the MCB copper trace that cut connection to the +5V ISO regulator that inhibits ON/OFF function.
We found a shorted back diode in the Enode Board
Both the MCB damaged copper trace was fix as well as the Enode shorted diode was remove. A missing back diode inthe Main Contactor 24V DC coil added. System was down for a year.

Address

Mandaluyong
1550

Alerts

Be the first to know and let us send you an email when Geramil Technology Services posts news and promotions. Your email address will not be used for any other purpose, and you can unsubscribe at any time.

Contact The Practice

Send a message to Geramil Technology Services:

Share

Share on Facebook Share on Twitter Share on LinkedIn
Share on Pinterest Share on Reddit Share via Email
Share on WhatsApp Share on Instagram Share on Telegram

Category